| 8 | _binary | 2024-11-11 00:00:00.000000 | 3 | 1 | After evaluation 34 cables failed the leak test and 4 cables passed. The cables that passed were 001., 003, 019, 021. It was observed that there was a cap on the bundle that would allow air to travel through during the leak test. It was also observed that there is a wire mark from the EDM process from the suppliers for both the crescent bushing and the distal shell. CAPA 000413 has been opened. | 3 | 38.0 | 1 | _binary | 8 | _binary | 40 | 1 |