{"database": "fishbowl_archive", "table": "rmaitem", "rows": [[8, "_binary \u0000", "2024-11-11 00:00:00.000000", 3, 1, "After evaluation 34 cables failed the leak test and 4 cables passed.  The cables that passed were 001., 003, 019, 021.  It was observed that there was a cap on the bundle that would allow air to travel through during the leak test.  It was also observed that there is a wire mark from the EDM process from the suppliers for both the crescent bushing and the distal shell.\n\nCAPA 000413 has been opened.", 3, null, 38.0, 1, "_binary \u0000", 8, "_binary \u0000", 40, 1, null, ""]], "columns": ["id", "crossShipFlag", "dateResolved", "issueId", "lineNum", "note", "productId", "productSubId", "qty", "resolutionId", "returnFlag", "rmaId", "shipFlag", "typeId", "uomId", "vendorId", "vendorRMA"], "primary_keys": ["id"], "primary_key_values": ["8"], "units": {}, "query_ms": 2.959254023153335}