id crossShipFlag dateResolved issueId lineNum note productId productSubId qty resolutionId returnFlag rmaId shipFlag typeId uomId vendorId vendorRMA
44 _binary   3 1 Disposition: After evaluation of the 17 cables returned off RGA 2496 Qty of 17 we have confirmed that 16 pcs failed the leak test and 1 pc passed. Immediate Correction: We will rework the 16 pcs and return the 1 cable that is good with no rework performed. Proposed Preventive Action: During review of the validation report, it was confirmed that the cables were validated to withstand a single pressure test at 3.9 PSI. However, testing performed at Innofiber and the customer's facility includes: A.) A single test at 8 PSI which is more than double the validted pressure. B.) Multiple test at 4 PSI, which also exceed the validated conditions. Innofiber considers testing at 8 PSI , for 30 seconds to be destructive in nature and this has been communicated to ISI . As such, no CAPA will be opened for this issue. While the conditions fall outside the scope of warranty coverage, Innofiber is willing to evaluate and repare cables when possible. Please note: No- nonconformance reports will be accepted for cables returned due to leaks resulting from over-pressure testing. Repairs will be performed at a cost quoted by our sales team, based on the extent of damage and feasibility of repair. 3   17.0   _binary 26 _binary  40 1    